Chapter 10

Magnetic mobile microrobots for mechanobiology and automated biomanipulation

Wuming Jing; Sagar Chowdhury; David Cappelleri    Purdue University, West Lafayette, IN, United States

Abstract

This chapter presents our recent efforts on developing a micro-force sensing mobile microrobot (μFSMM). The design consists of a planar, vision based micro-force sensor end-effector attached to a magnetic microrobot body. The body is made of chemically etched nickel and is driven by an exterior magnetic field. With a known stiffness, the manipulation forces can be determined from observing the deformation of the end-effector through a camera attached to an optical microscope. After analyzing and calibrating the stiffness of a micromachined prototype, the mobility and in-situ force sensing capabilities are verified through force controlled manipulation tests. With our current experimental testbed, this micro-scale μFSMM is able to translate with the speed up to 3 mm/sImage in an oil environment. The calibrated stiffness of the micro-force sensor end-effector is on the order of 102 N/mImage. The force sensing resolution is approximately 100 nNImage.

Keywords

Micro-force sensing; Magnetic micro/nanorobots

10.1 Introduction

Mechanical forces at the micro-scale have been recognized as critical factors determining various biological functions. The study of cell or tissue mechanics is critical to explain problems in physiology and disease development. In fact, the fields of mechanotransuction [1] and mechanobiology [2] are devoted to these areas. Knowledge of these forces is also needed for safe biomanipulation [3]. Current tools for cell manipulation and sensing micro-forces all require complex tethered hardware that clutters the workspace. In addition, these tethered settings limit the force sensing probe from certain enclosed in-vivo working environments. Therefore, untethered force sensing techniques such as computer vision based sensing is favorable in these bio-applications. On the other hand, in order to take the advantage of an untethered working mode, we also need an untethered device to move the force-sensing end-effector to perform manipulation tasks, such as manipulating a single cell. At the same time, submillimeter magnetic mobile microrobots have emerged as next generation robotic manipulators at the micro-scale. One of their most attractive features is the untethered actuation, which is since there are no power units small enough to carry on-board. Therefore, we are interested in integrating a magnetic mobile microrobot with a vision based micro-force sensing end-effector. As a result, a micro-force sensing mobile microrobot (μFSMM) has been developed (Fig. 10.1). It can serve as an ideal option to perform daily bio-manipulation tasks with low overhead.

Image
Figure 10.1 Road map of micro-force sensing mobile microrobot (μFSMM) development.

This micro-force sensing mobile microrobot system inherits the magnetic microrobot system and can easily be integrated into standard bio-testbeds. As shown in Fig. 10.2, the μFSMM incorporates a two-dimensional (2D) vision based micro-force sensor with a micro-sized magnetic body. Thus, the untethered device can be driven as a mobile microrobot by an external magnetic field that is produced by a electromagnetic coil system. When driving this μFSMM to manipulate a single cell or micro-object, the micro-force sensor will deflect a certain amount, which can be detected by the camera system underneath the workspace. With this deflection feedback and known planar stiffness of the micro-force sensor, one can derive the amount of in-situ force acting on the robot. Therefore, force controlled bio-manipulation and cell or tissue characterization is possible with these μFSMMs.

Image
Figure 10.2 Concept of operations of a diagram of micro-force sensing mobile microrobot (μFSMM) system [4].

This chapter will first briefly review the state-of-the-art of the micro-force sensing and mobile magnetic microrobot technologies. Then the design, analysis, and fabrication of the micro-scale μFSMM will be introduced in further detail. Afterwards, experimental tests will be demonstrated to verify the μFSMM's mobility and capability of performing force controlled manipulation tasks.

10.2 State-of-the-art: micro-force sensing, mobile magnetic microrobot

10.2.1 Micro-force sensing

There has been mounting evidence indicating that mechanical forces can play a critical role in physiology and disease development [5]. Micro-manipulation and micro-force sensing techniques have been investigated to facilitate this exploration of mechanobiology.

10.2.1.1 Micro-manipulation of biology specimens

The cells, nucleus, and other biological specimens that have been manipulated by the engineering methods include (i) mechanical, (ii) optical, and (iii) magnetic means. For mechanical means, one classical technique is micropipette aspiration. The micropipette can apply a low negative pressure to deform and elongate a portion of a cell into it. Based on the deformation, various continuum models [68] can be built for quantifying the cell's mechanical properties, such as elasticity and viscosity. One of typical optical methods is laser trapping, also known as optical trap or laser tweezer. The laser beam can create a potential well to trap micro-particles within a defined region. Thus, the cell can be precisely manipulated when attached to the micro-particle being trapped by the laser tweezer. The stretching force in between the micro-particle and cell is proportional to the applied laser power, where the force generated is approximately 0.1–1 nN. Applying the principle of laser trapping, various live entities have been investigated, including virus and bacteria [9], red blood cells [10], natural killer cells [11], and outer hair cells [12]. These laser tweezers have further been refined and innovated through optoelectronic tweezers [13], laser-tracking micro-rheology (LTM) [14], and surface plasmon resonance excited by polarized light [15]. Similar to the micro-particles in laser tweezer, magnetic micro-beads can serve as the handle for magnetic tweezer [16], where the manipulation force equals to the acting magnetic force. These magnetic micro-beads can be ligand-coated [17]. They can work in magnetic gradient [18] and magnetic twisting cytometry (MTC) [19].

10.2.1.2 Micro-force sensing

Beyond the more traditional micro-manipulation methods mentioned in the previous section, the progression of microelectromechanical system (MEMS) technology is significantly enhancing our capabilities in mechanical characterization of the biological specimens. For example, Galbraith et al. [20] measured traction forces generated by fibroblasts using a micro-fabricated device. Also with micro-fabricated force sensor, Yang and Saif et al. investigated the responses of adherent fibroblasts to stretching forces [21] and the role of mechanical tension in neurotransmission [22]. The representative working principles of MEMS micro-force sensors include piezoresistive sensors [23], strain gauges [24], and comb-capacitance structures [25,26]. They are able to sense the micro-forces in magnitudes of less than 1 mNImage. These MEMS devices are not convenient for real biological applications in fluid environments due to their tethered complex electronics. Moreover, these delicate MEMS devices are vulnerable to mechanical impact and thus, susceptible to failure.

Another family of micro- and nano-force sensors takes advantage of the sensing functions of an atomic force microscope (AFM) [27] or scanning electron microscope (SEM) [28]. For example, a vertical micro-cantilever in AFM can be operated to scan and deform a cell. The deflection of the AFM tip can be recorded to generate a map of stiffness across the cell's surface. The cellular and molecular characterization using this method has been reviewed in [29,30]. These customized modules in AFM or SEM are able to detect forces down to nN or even pN levels. However, these micro- and nano-force sensors require even more complex and larger auxiliary drive equipment, compared to the micro-force sensor based on MEMS devices. This creates a difficult test setup to replicate and use in a biology lab.

In contrast, vision-based force sensing can overcome many of the drawbacks mentioned above associated with MEMS and AFM/SEM micro-force sensors. As shown in Fig. 10.3, a stiffness calibrated elastic structure is the critical component of vision based force sensing. When the force F acts on the sensing probe, i.e., when pushing an target “cell”, the probe tip will deflect a certain amount Δ. Based on the known stiffness K of the elastic structure, we can determine the force F from

F=KΔ.

Image (10.1)

Image
Figure 10.3 The working principle of vision based force sensing.

In recent studies, the elastic object of low stiffness has been made of silicone elastomers with low Young's modulus and high failure strain. For example, a micro beam made of Polydimethylsiloxane (PDMS) has been used to sense one dimensional nano-Newton level forces for single cell studies [31]. Cappelleri et al. [32] have also used PDMS to develop a 2D vision based micro-force sensor for microrobotic and micromanipulation tasks. Park et al. [33] even made 3D structure out of PDMS for cell-polymer hybrid systems. Besides, micro-post arrays have also been micro-fabricated from silicone elastomer to measure forces exerted by single adhesion sites of a cell [34]. The image processing techniques for the deformation of micro-post arrays have been improved in [35,36]. These vision based micro-force sensors are able to provide force information feedback wirelessly. However, they cannot move to perform the manipulation task unless they are tethered to a motion module. Therefore, an untethered mobile micro-scale robot will be an ideal candidate to incorporate with the vision-based sensors perform wireless force controlled manipulation tasks.

10.2.2 Mobile magnetic microrobot

For micro-scale mobile microrobots (i.e., robots with largest dimension of <1 mm), magnetic actuation is one of most widely applied working mechanisms due to its intrinsic feature of wireless power delivery. Based on magnetic principles, micro-scale magnetic microrobots can be directly driven by magnetic field gradients [37,38]. Other alternative magnetic field commands have also been implemented to enhance the microrobot's mobility to operate in complex working environments. Some representative working mechanisms include oscillating [39], rocking (stick–slip) [40], rolling [41,42], and tumbling [43,44] locomotions. However, these magnetic microrobots are all micro-sized magnets in essence, and no more functionality has been typically demonstrated other than mobility.

10.3 Micro-force sensing mobile microrobot

The micro-force sensing mobile microrobot (μFSMM) arms the magnetic microrobot with vision based micro-force sensor in order to realize a new class of wireless micro-force sensing robots. On the one hand, μFSMM liberates the vision based micro-force sensor from a fixed manipulator. Therefore, it is able to manipulate a cell and derive the micro-force information during the process in an untethered way. On the other hand, μFSMM arms the untethered vision based micro-force sensor on the wireless magnetic mobile microrobot. In this way, magnetic microrobot is conferred with wireless force sensing abilities. Adding this vision based micro-force sensor yields a new powerful capability without magnetic mobile microrobots. Therefore, the μFSMM is an ideal micro-scale robotic end-effector to perform daily biological manipulation tasks, and has a simple setup and low cost.

In the following section, the customized magnetic microrobot system for μFSMMs will be introduced first. It can be easily fit onto the standard inverted optical microscope systems used for biological tests. Afterwards, the design, analysis, fabrication and experimental test of the micro-scale μFSMM prototypes will be illustrated in further details.

10.3.1 Microscope compatible magnetic coil testbed for μFSMM

In order to power the μFSMM, a mobile magnetic microrobot essentially, a magnetic coil testbed is required (Fig. 10.4). This electromagnetic coil testbed is newly designed and built for biological testing. As shown in Fig. 10.4, the coil system consists of four electromagnetic coils mounted horizontally, encompassing the workspace, that can be observed from either an overhead or underneath camera position. The footprint of the coil system is approximately 160 mm×160 mmImage. It can hold a 35 mmImage diameter Petri dish with cultured cells, in which the μFSMM force sensing microrobot can also reside and be actuated by the magnetic field. The magnetic field is commanded by the power/driver unit. It consists of power supply, motor controller as driver unit for the coil, and a microcontroller board that takes commands from a PC. This power unit can produce as much as approximately 0.5 Tm1Image within the workspace. The magnetic field is modeled in COMSOL software (http://www.comsol.com) and verified with measurements from a Gauss meter.

Image
Figure 10.4 Microscope compatible magnetic coil testbed. (A) Overview of the coil testbed on an inverted microscope: (1) CMOS camera; (2) Electromagnetic coils; (3) Drive electronics unit. (B) Close-up views of the testbed: (4) Workspace, (5) Inverted microscope lens. (C) Evaluation of magnetic field with COMSOL. (D) Distribution of magnetic field in the workspace for an input current I = 2 A through one coil.

10.3.2 Design of μFSMMs

The μFSMM design consists of two primary sections (Fig. 10.5). The upper part of the frame is a vision based force sensor. The lower part of the frame consists of a micro magnetic body (hatched area) that functions as a mobile microrobot. The frame is made of silicon for mechanical strength. The joints are made of polydimethylsiloxane (PDMS) for low planar stiffness for the micro-force sensor.

Image
Figure 10.5 Diagram of the μFSMM design.

For manipulations of biological cells and mechanotransduction studies, the typical manipulation forces are in the low μN level and smaller. Thus, μFSMM microrobot should be able to: (i) exert manipulation force on this level; and (ii) sense the corresponding acting force. Moreover, the smaller footprint size of the microrobot, the easier it will be for it to access crowded narrow regions of the workspace. Therefore, the footprint of μFSMM should ideally be at the micro-scale (sub-mm). The detailed design specifications are described in the following subsections.

10.3.2.1 Design specifications

For design goal (i), the force that μFSMM can exert onto cells, it directly depends on the magnetic microrobot body because the acting magnetic force is the source of the moving momentum and also the manipulation force. The acting magnetic force, FmImage, can be evaluated as

Fm=V(MB)

Image (10.2)

where V is the magnetic volume, M is its magnetization, and B indicates the external magnetic field strength. The magnetic body is a micro-sized nickel part. The volume V (width × height × thickness) is w×h×t=300 μm×450 μm×76 μm=1.0×1011 m3Image. The magnetization M is set for simplicity as constant value of 3.5×105 Am1Image, which is based on the property of 99%Image pure annealed nickel in a magnetic field of 2 mTImage [45]. The magnetic field is produced by the electromagnetic coil testbed shown in Fig. 10.4. The field gradient can be greater than 0.5 Tm1Image within the workspace. Therefore, a representative magnetic force, FmImage, acting on the current μFSMM can be as large as Fm=V(MB)1.75 μNImage.

On the other hand, for the design goal (ii), this amount of force needs to be detected by the on-board micro-force sensor as well. For vision based force sensing, the sensed force, F, is determined by the known stiffness, K, of the elastic sensor structure and its detected deflection Δ due to the applied force. This is expressed as Eq. (10.1), namely, F=KΔImage. Under this applied force, the micro-force sensor needs to deflect larger than the minimum displacement, ΔminImage, that can be observed by the vision system. The ΔminImage of a vision system can be calculated as:

Δmin=2×Field of ViewCamera Sensor Resolution.

Image (10.3)

For our current overhead vision system, the field of view, camera sensor resolution, and other parameters are listed in Table 10.1. The vision system consists of a CMOS camera (FL3-U3-14E4C-C, http://www.ptgrey.com) and an adjustable 2.510XImage zoom imaging lens (VZM 1000i, http://www.edmundoptics.com). As shown in Table 10.1, the micro-force sensor must deflect more than Δmin=6.0 μmImage in order to be detected.

Table 10.1

Specifications of the vision system

Width (X) Height (Y) Unit Notes
Sensor size 7.18 5.32 mm Sensor format: 1/1.8”
Number of pixels 1280 1024 At 60 fps
Field of view 2.87 2.87 mm Lens at 2.5X
Observable size, ΔminImage 4.5 5.6 μm From Eq. (10.3)

Image

Therefore, the micro-force sensor should be no stiffer than the maximum stiffness, KmaxImage, when the applied force F results in the minimum observable displacement ΔminImage. It is obvious that F depends on the magnetic force FmImage, which varies with the magnetic microrobot body and the applied external magnetic field. ΔminImage varies with the available hardware of the vision system. For a conservative estimate as the initial reference for μFSMM design, let F=Fm/3Image. Then the required stiffness of the micro-force sensor KKmax=F/Δmin=Fm/3(Δmin)0.1 Nm1Image.

10.3.2.2 Design of micro-force sensor end-effector

After the stiffness requirement is determined, the detailed design of the micro-force sensor end-effector shown in Fig. 10.5 can be done. In order to meet the low stiffness requirement, the material of elastic section is selected as a silicone elastomer, Polydimethylsiloxane (PDMS), whose Young's modulus is no more than several MPa. The rest of the frame section and end-effector is made from silicon, which is selected for its mechanical strength to avoid out-of-plane deflection and easy workability with MEMS fabrication processes. Moreover, a micro-scale spring-type compliant structure is chosen to yield low planar stiffness. The geometry design is based on our previous studies on the meso-scale vision based force sensor [32]. Here, we introduce the lower stiffness spring structure design within a micro-scale envelope. In addition, the shuttle probe is customized into various shaped end-effectors to facilitate different manipulation processes (i.e., such as cell pushing).

The stiffness of the micro-force sensor design is analyzed with finite element analysis in COMSOL (http://www.comsol.com). A series of designs with tuned dimensions are modeled. The results indicate that four geometric parameters can have different influences on device stiffness. They are summarized in Table 10.2 (for the data in the table, only one design parameter was varied at a time). Geometric parameters impacting the structure's stiffness include: txy, l1, l2Image, and d, as shown in Fig. 10.5, where txyImage is the beam width, l1, l2Image are the beam lengths, and d indicates the gap between the beams. As shown in Table 10.2, the beam width, txyImage, and the gap between beams, d, can tune the device stiffness significantly, while the beam lengths, l1, l2Image, impact the stiffness in a minor degree. Note that the thickness of the device, tzImage, also impacts the stiffness proportionally. Thus, it is set as 50 μmImage during the analysis.

Table 10.2

Impacts of geometric parameters on the stiffness of micro-force sensor

Impact Beam width (txyImage) ↑ Beam length (l1Image & l2Image) ↑ Gap between beams (d) ↑
Device stiffness in X axis
Device stiffness in Y axis

Image

Note: This table illustrates the response of micro-force sensor stiffness with increases in the geometric parameters. “↑ / ↓” means slightly increase/decrease: when geometric parameter varies >50%Image, stiffness changes <5%Image. “⇑ / ⇓” means significant increase/decrease: when geometric parameter varies <15%Image, stiffness changes >50%Image.

The selected geometric parameters for the micro-force sensor designs are summarized in Table 10.3. The different combinations of these geometric parameters result in 32 different μFSMM designs (P1–P32). Their footprints are between 595 μm×765 μmImage to 890 μm×875 μmImage, which meet the micro-scale dimension requirement. The FEA models show the largest stiffness is approximately 0.05 Nm1Image, which is well less than 0.1 Nm1Image, the estimated stiffness requirement.

Table 10.3

Geometric parameters of the micro-force sensor

Prototype # (1–32) Beam width txyImage (μm) Gaps between beams d (μm) Beam lengths l1Image & l2Image (μm)
1 – 4 10 45 (75, 75) (50, 50) (50, 75) (25, 50)
5 – 8 60
9 – 12 15 45
13 – 16 60
17 – 20 20 45
21 – 24 60
25 – 28 25 45
29 – 32 60

Image

Note: The thickness of the device tz=50 μmImage.

10.3.3 Prototype fabrication of μFSMM

The micro-scale μFSMM prototypes have been fabricated with a two phase process (Fig. 10.6). Phase I manufactures the micro-force sensor and silicon frame parts with MEMS processes; this phase is based on the process using silicon on insulator (SOI) wafer described in [46]. Phase II subsequently assembles the released micro-force sensor with micro-sized magnetic body. Thereafter, the total device can serve as the μFSMM. Unlike the SOI wafer, the fabrication of the μFSMM starts with a 300 μmImage-thick bare silicon wafer without an oxide layer. This is done in order to customize the thickness of device layer, which is proportionally related to the stiffness of the micro-force sensor.

Image
Figure 10.6 Fabrication process of microscale μFSMM prototypes. (A) Phase I – fabrication of compliant micro-force sensor and rigid frame components. (B) Phase II – fabrication and assembly of magnetic body to frame. (C) Assembled prototypes.

In step 1 of phase I, positive photoresist AZ 1518Image (http://www.microchemicals.com) is coated (2 μmImage) and patterned with photolithography. This thin patterned layer serves as the etching mask of the following deep reactive ion etch (DRIE) to etch out the mold for the PDMS joints. This step essentially etches to the thickness of the micro-force sensor and determines the device thickness (50 μmImage here for the initial prototypes). In step 2, the PDMS (Sylgard 184, 10:1 ratio of base to curing agent, http://www.dowcorning.com) is prepared, coated and cured at 60°CImage overnight. The sample is then bathed in acetone to remove the redundant PDMS layer due to the sacrificial photoresist layer in the previous step. The PDMS residues are further cleaned by another wet chemical etching bath (nmethylpyrrolidone:tetrabutylammonium fluoride in 3:1 ratio) for 1 minImage. In step 3, another thin AZ 1518Image photoresist layer is patterned to define the rest of the device frame. It is also etched to the same depth as in step 1 using a DIRE process. In step 4, another DRIE process is performed from the back-side windows that are aligned with front-side structure and patterned with a thick photoresist layer AZ 9260Image (8 μmImage). This through etching will automatically release the front-side structures. Therefore, at the end of phase I, the frame part embedded with the micro-force sensor is constructed.

In phase II, the released independent micro-force sensors are assembled with a micro sized nickel part. The nickel part is manufactured by chemical etching (http://www.fotofab.com, step 5). The assembly is done by manually gluing (step 6) and then released by physical cutting. The final assembly serves as the wireless micro-force sensing mobile microrobot (μFSMM, Fig. 10.6(C)).

10.3.4 Experiments

Experimental tests have been conducted to verify the functionality of manufactured μFSMM. The stiffness of the micro-force sensor module is calibrated in a customized setup. After the mobility tested, force controlled manipulation tests on cell analogs have been realized using the micro-scale μFSMM prototypes. This demonstrates that μFSMM simultaneously possesses mobility and capability of wireless micro-force sensing.

10.3.4.1 Prototype calibration

At first, the stiffness of the micro-force sensor on the μFSMM is experimentally calibrated using an electrostatic MEMS micro-force sensing probe (FT-S1000, resolution 0.05 μN @ 10 HzImage, http://www.femtotools.com). As shown in Fig. 10.7, while the micro-force sensor of μFSMM is fixed, the micro-force sensing probe FT-S1000 is mounted on a micromanipulator system (MPC-200, http://www.sutter.com). This micromanipulator arm can translate along XYZImage directions independently with a resolution of 1 μmImage. Thus, the micro-force sensing probe can be leveled and flushed with the micro-force sensor end-effector. During the calibration tests, the traveled distance of the force sensing probe is recorded in real-time. This corresponds to the deflection of the calibrating micro-force sensor, Δ. Therefore, the stiffness of the fabricated prototype, K, can be derived based on Eq. (10.1), namely, K=F/ΔImage, where F is the blocking force read out from the micro-force sensing probe. These stiffness calibrations also show that the vision based micro-force sensor on the μFSMM is able to sense the same level force that can be read by the MEMS device based micro-force sensing probe with tethered electronics.

Image
Figure 10.7 The stiffness calibration test setup for the micro-force sensor end-effector of the μFSMM [47].

The stiffness calibration results of the #132Image micro-force sensor designs are summarized in Fig. 10.8. In the measurement testing, the micro-force sensor tip is pushed with a deflection of 20 μmImage to 80 μmImage. Comparing with the FEA results, the stiffness of the actual prototypes are all stiffer, with values at most approximately 25%Image stiffer. This can be explained by the variance of: (i) geometric dimensions; and (ii) material stiffness properties for the prototypes.

Image
Figure 10.8 Comparison between the stiffness calibration results of the micro-force sensor end-effectors and their FEA results.

For reason (i), the geometric dimension of the fabricated prototype can be different from the nominal values. The planar geometric dimensions of the fabricated prototypes show a small deviation from nominal values (3 μmImage), which is essentially the resolution of photolithography process. This is a minor factor explaining the stiffness variance. The thickness of the device layer (tzImage) is also one of the geometric parameters that can impact the stiffness of micro-force sensor. Any extra thickness on the elastic structure can be interpreted as two spring elements bundled in parallel. Therefore, the thickness deviation (<5%Image) also cannot fully explain the increased stiffness (25%Image) of the μFSMM prototypes.

For reason (ii), the material property of the PDMS elastomer may be different from the theoretical values plugged in the FEA model. The stiffness of PDMS is a function of curing temperature and aging [48]. The Young's modulus can even reach to >2 MPaImage after a long curing under high temperatures (100CImage). This can account for the PDMS hardening in μFSMM prototypes as they are exposed to high temperature during MEMS fabrication processes. After the PDMS is cured, the wafer sample still needs two steps of photolithography using AZ 1518Image and AZ 9260Image, respectively. Both of them require soft bake at >100CImage before exposure. This excessive baking process will increase the Young's modulus of the final PDMS structure, which is unavoidable. To verify this hardening, a higher Young's modulus is plugged in the FEA model. At approximately 800 kPa1.1 MPaImage, the FEA will have the same result as the calibration results (discrepancy <5%Image).

Given this stiffness variance, the calibrated stiffness of the fabricated μFSMM prototypes (Fig. 10.8) are still well in the stiffness range required by our current vision and drive system (100×103 N/mImage). Limited work has been done in microrobot force sensing, including the efforts from Kawahara et al. [49]. Table 10.4 shows a comparison of primary specifications between the work in [49] and the calibrated micro-scale μFSMM prototypes. These μFSMM prototypes have shown clear advantage when compared with this other microrobot with force sensing capabilities.

Table 10.4

Comparison of specifications: μFSMM versus the force sensing microrobot sensing device in [49]

Force sensing microrobot in [49] μFSMM
Microrobot size ∼5 mm <1 mm
Micro-force sensor size ∼750 μm 500–800 μm
Micro-force sensor material Silicon Elastomer
Micro-force sensor stiffness 0 N/m <0.1 N/m
Force sensing range ∼0–5 mN ∼0–3 μN
Sensing resolution ∼80 μN ∼0.1 μN

10.3.4.2 Mobility

Mobility tests are performed with assembled μFSMM prototypes. The microrobot is submerged into an oil bath and rests on a silicon substrate during the tests. The fluid medium is silicone oil (A12728, http://www.alfa.com), whose dynamic viscosity ν is 40 c.s.t.Image Actuated by the external magnetic field in the workspace, the first series of tests to demonstrate mobility drive the microrobot in a linear translation (Fig. 10.9(A)). With a 10 mTImage field, the microrobot is able to move at approximately 2 mm/sImage with these specific settings. The second mobility test rotates the microrobot in place (Fig. 10.9(B)). The ability to adjust orientation is needed for cell manipulation to control the specific location and approach angle of the robot when approaching and manipulating the cell. In order to realize steady controllable rotation without translation, a magnetic field with a low intensity is applied, less than 4 mTImage. Under the applied field signal, the microrobot rotates in place with the end-effector tip of the microrobot following a circular trajectory, with a deviation error less than 20 μmImage.

Image
Figure 10.9 Mobility test of μFSMM. (A) Linear motion: linear velocities at increasing magnetic field intensity up to 10 mT. (B) Rotation: the trajectory of the tip of the microrobot during a 360 pure rotation. The actuating field (B) is 4 mT and the angular velocity (ω) is 0.42 rad s−1.

10.3.4.3 Manipulation on cell analog with micro-force feedback

The μFSMM was then tested to evaluate its ability to push a cell analog (micro-disc with diameter =200400 μmImage, thickness =50 μmImage) and simultaneously detect the acting force. The testing environment is the same as in the case of the mobility tests (oil bath, viscosity ν=40 c.s.t.Image, silicon substrate). Snapshots for one such test are shown in Fig. 10.10. The μFSMM microrobot is first driven to the target “cell”. After a short pause, the microrobot is driven to push the cell. During the push, the micro-force sensor end-effector deflects up to Δ30 μmImage. Eventually, the “cell” is nudged and displaces an amount approximately equal to the deflection, Δ, and the micro-force sensor returns to its original shape (Δ=0Image). This particular pushing test is performed with μFSMM prototype #9 from Table 10.3. According to the calibrated stiffness (Fig. 10.8), the stiffness of its micro-force sensor end-effector in the Y direction, KY0.01 N/mImage. Therefore, the maximum force during this pushing action is F=KY×Δ0.3 μNImage. The behavior of the μFSMM during this test is shown in Fig. 10.11.

Image
Figure 10.10 Pushing manipulation test of μFSMM (applied magnetic field B ≈ 5 mT). (A) Original position. (B) Onset of pushing after approaching the target “cell”. (C) The instant during pushing when the micro-force sensor has the maximum deflection (≈30 μm). Note: O is the tip of end-effector, R is reference point on the microbot body. The deflection of the micro-force sensor is determined by comparing the difference of the length of the vectors ORImage and ORImage.
Image
Figure 10.11 μFSMM characteristics during pushing manipulation. The lower plot shows the displacement of the microrobot body with time. The upper plot indicates the deflection and sensed force of the micro-force sensor end-effector with time, during the pushing process.

In real cell manipulation tests, the desired micro-force information is the amount of force, F0Image, that is purely applied on the cell. In the above pushing test, the sensed force F by the μFSMM not only consists of F0Image, but actually

F=F0+f+FD

Image (10.4)

where F0Image indicates the force acted on the “cell”, f is the friction force against substrate, and FDImage represents the fluid drag during the motion of microrobot. Therefore, both the friction f and fluid drag FDImage also need to be evaluated before the actual force F0Image on cell is revealed.

Based on Coulomb's model, the friction f=μGImage, where G is the gravity of the microrobot prototype and μ is the friction coefficient. For estimation, G is calculated for a 600×500×50Image (μm3Image) silicon block, resulting in G0.34 μNImage. The friction coefficient μ is plugged in as 0.03, similar scenario here as in [50]. Therefore, the friction force f1×102 μNImage.

For the fluid drag, FDImage, the Reynolds number Re is evaluated first, which is defined as

Re=ρvl/ν

Image (10.5)

where ρ is the mass density of the oil bath (960 kg/m3Image); v is the fluid velocity, which can be set as the microrobot's speed during the pushing action (50 μm/sImage); l is the characteristic length, which is the length of the microrobot (750 μmImage); and ν is the dynamic viscosity of the silicone oil (40 c.s.t.=0.04 PasImage). Thus, for this pushing test in Fig. 10.10, Re9×104Image, which is well within the laminar flow domain. In order to further evaluate the fluid drag force, FDImage, the laminar flow model of μFSMM is built in COMSOL software (Fig. 10.12). Based on the pressure results (Fig. 10.12(B)), the fluid drag FDImage can be assessed by the surface integration of the Y component of the pressure on all the faces of the microrobot. This allows us to estimate the fluid drag, FDImage, on the direction of microrobot movement (Y axis) as approximately 8×103 μNImage.

Image
Figure 10.12 The FEA model in COMSOL for the evaluation of fluid drag force FD. (A) The velocity field. (B) The pressure distribution on the surface of μFSMM microrobot.

Summing up the friction, f, and fluid drag, FDImage, accounts only for approximately 5%Image of the sensed force, F in Eq. (10.4). Therefore, the micro-force sensor on the μFSMM is able to provide the in-situ feedback of the force acting on the “cell”, which is on the order of 0.1 μNImage, with an accuracy of 95%Image.

10.3.4.4 Path planned manipulation with force control

After verifying the mobility and micro-force sensing function of μFSMM, a comprehensive manipulation test is performed with both path planned motion and force controlled manipulation (Fig. 10.13). The working environment is the same as previous experimental tests.

Image
Figure 10.13 Path planned manipulation using μFSMM with force control. (A) Original position. (B) Off-line path planning in MATLAB for navigating the microrobot. (C) Automated navigation to approach the first manipulation object. (D) Manual control to manipulate the first object “cell” to the target location. (E) Manual control to manipulate the second object “cell” to the target location. (F) Automated (backwards) navigation to the original position.

As shown in Fig. 10.13(A), a μFSMM microrobot is placed preparing to push two object “cell” analogs into target position marked by an “L” bay. Then, the path navigated to the first “cell” is planned in MATLAB using the AImage algorithm [51] (Fig. 10.13(B)). Thereafter, the microrobot executes the planned route approaching to the first “cell” analog (Fig. 10.13(C)). After approaching the manipulation object, the μFSMM microrobot is manually controlled pushing the “cell” into the target zone. The reason for conducting the manual control during manipulation is the non-prehensile essence of the micro-manipulation, which is difficult to automate and is not the focus of this particular work. Afterwards, the second objective “cell” analog is pushed and stacked into the target zone. After finishing all the manipulation tasks, the microrobot is automatically driven back to the original position. In this manipulation test, the applied forces can be controlled based on the force feedback from the vision based micro-force sensor end-effector. When manually controlling the μFSMM microrobot to push the “cell”, the manipulation can be stopped when the applied force is larger than the maximum allowed amount.

10.4 Concluding remarks

This chapter has introduced our recent work on developing the micro-force sensing mobile microrobot (μFSMM) design at micro-scale. The untethered working mode of μFSMM is an ideal option for routine bio-manipulation tests. The realization of μFSMM merges the mobile magnetic microrobot and vision based micro-force sensing techniques. The work in this chapter has verified this μFSMM concept with design, prototype fabrication, and experimental manipulation tests with “cell” analogs. In order to adopt this μFSMM design to real biological applications, further work needs to be done in the following areas.

The current micro-scale μFSMM prototype needs to be further scaled down, since it's still larger than most of the cell species. This will allow for easier interaction with real cell or tissue samples. What's more important is that the micro-force sensing range needs to be further refined to the nN range. This is the maximum amount of force that many cells can withstand. In addition, it is still very difficult to automate the active manipulation of the micro-scale objects using a mobile microrobot. The non-prehensile characteristics make it even more different in addition to the unique forces present at the micro-scale. However, automated force-controlled micro-manipulation is the ultimate goal to conduct reliable, safe, and high throughput biological experiments.

References

[1] D.E. Jaalouk, J. Lammerding, Mechanotransduction gone awry, Nature Reviews Molecular Cell Biology 2009;10:63–73.

[2] P. Hersen, B. Ladoux, Biophysics: push it, pull it, Nature 2011;470:340–341.

[3] J.P. Desai, A. Pillarisetti, A.D. Brooks, Engineering approaches to biomanipulation, Annual Review of Biomedical Engineering 2007;9:35–53.

[4] W. Jing, D. Cappelleri, Incorporating in-situ force sensing capabilities in a magnetic microrobot, In: Proc. of IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS). 2014.

[5] D.H. Kim, P.K. Wong, J. Park, A. Levchenko, Y. Sun, Microengineered platforms for cell mechanobiology, Annual Review of Biomedical Engineering 2009;11:203–233.

[6] M. Sato, D.P. Theret, L.T. Wheeler, N. Ohshima, R.M. Nerem, Application of the micropipette technique to the measurement of cultured porcine aortic endothelial cell viscoelastic properties, Journal of Biomechanical Engineering 1990;112:263–268.

[7] D. Needham, R.M. Hochmuth, Rapid flow of passive neutrophils into a 4 μm pipet and measurement of cytoplasmic viscosity, Journal of Biomechanical Engineering 1990;112:269–276.

[8] D.E. Discher, D.H. Boal, S.K. Boey, Simulations of the erythrocyte cytoskeleton at large deformation. II. Micropipette aspiration, Biophysical Journal 1998;75:1584–1597.

[9] A. Ashkin, J. Dziedzic, Optical trapping and manipulation of viruses and bacteria, Science 1987;235:1517–1520.

[10] M. Dao, C. Lim, S. Suresh, Mechanics of the human red blood cell deformed by optical tweezers, Journal of the Mechanics and Physics of Solids 2003;51:2259–2280.

[11] S. Seeger, S. Monajembashi, K. Hutter, G. Futterman, J. Wolfrum, J. Greulich, Application of laser optical tweezers in immunology and molecular-genetics, Cytometry 1991;12:497–504.

[12] D. Murdock, S. Ermilov, A. Spector, A. Popel, W. Brownell, B. Anavri, Effects of chlorpromazine on mechanical properties of the outer hair cell plasma membrane, Biophysical Journal 1989;89:4090–4095.

[13] P. Chiou, A. Ohta, M. Wu, Massively parallel manipulation of single cells and microparticles using optical images, Nature 2005;436:370–372.

[14] S. Yamada, D. Wirtz, S.C. Kuo, Mechanics of living cells measured by laser tracking microrheology, Biophysical Journal 2000;78:1736–1747.

[15] X.Y. Miao, L.Y. Lin, Trapping and manipulation of biological particles through a plasmonic platform, IEEE Journal of Selected Topics in Quantum Electronics 2007;13:1655–1662.

[16] C. Gosse, V. Croquette, Magnetic tweezers: micromanipulation and force measurement at the molecular level, Biophysical Journal 2002;82:3314–3329.

[17] N. Wang, J. Butler, D. Ingber, Mechanotransduction across the cell-surface and through the cytoskeleton, Science 1993;260:1124–1127.

[18] A. Bausch, F. Ziemann, A. Boulbitch, K. Jacobson, E. Sackmann, Local measurements of viscoelastic parameters of adherent cell surfaces by magnetic bead microrheometry, Biophysical Journal 1998;75:2038–2049.

[19] N. Wang, D. Ingber, Probing transmembrane mechanical coupling and cytomechanics using magnetic twisting cytometry, Biochemistry and Cell Biology 1995;73:327–335.

[20] C.G. Galbraith, M.P. Sheetz, A micromachined device provides a new bend on fibroblast traction forces, Proceedings of the National Academy of Sciences of the United States of America 1997;94:9114–9118.

[21] S. Yang, T. Saif, Reversible and repeatable linear local cell force response under large stretches, Experimental Cell Research 2005;305:42–50.

[22] S. Yang, S. Siechen, J. Sung, A. Chiba, T. Saif, MEMS based sensors to explore the role of tension in axons for neuro-transmission, In: 21st IEEE Int. Conf. MEMS. 2008.

[23] J. Wei, M. Porta, M. Tichem, U. Staufer, P.M. Sarro, Integrated piezoresistive force and position detection sensors for micro-handling applications, Journal of Microelectromechanical Systems 2013;22:1310–1326.

[24] A. Hoover, R. Fearing, Rapidly prototyped orthotweezers for automated microassembly, In: Proc. of IEEE International Conference on Robotics and Automation (ICRA). 2007.

[25] Y. Sun, B.J. Nelson, MEMS for cellular force measurements and molecular detection, International Journal of Information Acquisition 2004;1:23–32.

[26] F. Beyeler, S. Muntwyler, B.J. Nelson, A six-axis MEMS force-torque sensor with micro-Newton and nano-Newtonmeter resolution, Journal of Microelectromechanical Systems 2009;18:433–441.

[27] E.M. Puchner, H.E. Gaub, Force and function: probing proteins with AFM-based force spectroscopy, Current Opinion in Structural Biology 2009;19:605–614.

[28] M.R. Ahmad, M. Nakajima, M. Kojima, S. Kojima, M. Homma, T. Fukuda, Instantaneous and quantitative single cells viability determination using dual nanoprobe inside ESEM, IEEE Transactions on Nanotechnology 2012;11:298–306.

[29] M. Radmacher, Measuring the elastic properties of living cells by the atomic force microscope, Methods in Cell Biology 2002;68:67–90.

[30] A. Kedrov, H. Janovjak, K.T. Sapra, D.J. Muller, Deciphering molecular interactions of native membrane proteins by single-molecule force spectroscopy, Annual Review of Biophysics and Biomolecular Structure 2007;36:233–260.

[31] F. Sasoglu, A. Bohl, B. Layton, Design and microfabrication of a high-aspect ratio PDMS microbeam array for parallel nanonewton force measurement and protein printing, Journal of Micromechanics and Microengineering 2007;17:623–632.

[32] D.J. Cappelleri, G. Piazza, V. Kumar, A two dimensional vision-based force sensor for microrobotic applications, Sensors and Actuators A, Physical 2011;171:340–351.

[33] J. Park, J. Kim, D. Roh, S. Park, B. Kim, K. Chun, Fabrication of complex 3d polymer structures for cell-polymer hybrid systems, Journal of Micromechanics and Microengineering 2006;16:1614–1619.

[34] N.Q. Balaban, U.S. Schwarz, D. Riveline, P. Goichberg, G. Tzur, I. Sabanay, D. Mahalu, S. Safran, A. Bershadsky, L. Addadi, B. Geiger, Force and focal adhesion assembly: a close relationship studied using elastic micropatterned substrates, Nature Cell Biology 2001;3:466–472.

[35] C.A. Lemmon, N.J. Sniadecki, S.A. Ruiz, J.L. Tan, L.H. Romer, C.S. Chen, Shear force at the cell–matrix interface: enhanced analysis for microfabricated post array detectors, Mechanics & Chemistry of Biosystems 2005;2:1–16.

[36] B. Li, L. Xie, Z. Starr, Z. Yang, J. Lin, J. Wang, Development of micropost force sensor array with culture experiments for determination of cell traction forces, Cell Motility and the Cytoskeleton 2007;64:509–518.

[37] M. Kummer, J.J. Abbott, B.E. Kratochvil, R. Borer, A. Sengul, B.J. Nelson, OctoMag: an electromagnetic system for 5-DOF wireless micromanipulation, IEEE Transactions on Robotics 2010;26:1006–1017.

[38] S.N. Tabatabaei, J. Lapointe, S. Martel, Shrinkable hydrogel-based magnetic microrobots for interventions in the vascular network, Advanced Robotics 2011;25:1049–1067.

[39] H.-W. Tung, M. Maffioli, D.R. Frutiger, K.M. Sivaraman, S. Pané, B.J. Nelson, Polymer-based wireless resonant magnetic microrobots, IEEE Transactions on Robotics 2014;30:23–32.

[40] C. Pawashe, S. Floyd, M. Sitti, Modeling and experimental characterization of an untethered magnetic micro-robot, The International Journal of Robotics Research 2009;28:1077–1094.

[41] G.L. Jiang, Y.H. Guu, C.N. Lu, et al., Development of rolling magnetic microrobots, Journal of Micromechanics and Microengineering 2010;20, 085042.

[42] H.-W. Tung, K.E. Peyer, D.F. Sargent, B.J. Nelson, Noncontact manipulation using a transversely magnetized rolling robot, Applied Physics Letters 2013;103:114101.

[43] W. Jing, N. Pagano, D.J. Cappelleri, A novel micro-scale magnetic tumbling microrobot, Journal of Micro-Bio Robotics 2013;8:1–12.

[44] W. Jing, N. Pagano, D. Cappelleri, A tumbling magnetic microrobot with flexible operating modes, In: Proc. of IEEE International Conference on Robotics and Automation (ICRA). 2013.

[45] R.M. Bozorth, Ferromagnetism. Wiley–IEEE Press; 1993.

[46] A.P. Gerratt, I. Penskiy, S. Bergbreiter, SOI/elastomer process for energy storage and rapid release, Journal of Micromechanics and Microengineering 2010;20:104011.

[47] W. Jing, D. Cappelleri, Micro-force sensing mobile microrobots, In: Next-Generation Robotics II; and Machine Intelligence and Bio-inspired Computation: Theory and Applications IX. Proc. SPIE. 2015;vol. 9494 949405.

[48] D.T. Eddington, W.C. Crone, D.J. Beebe, Development of process protocols to fine tune polydimethylsiloxane material properties, In: 7th International Conference on Miniaturized Chemical and Biochemical Analysis Systems. 2003:1089–1092.

[49] T. Kawahara, M. Sugita, M. Hagiwara, F. Arai, H. Kawano, I.S. Ishikawa, A. Miyawaki, On-chip microrobot for investigating the response of aquatic microorganisms to mechanical stimulation, Lab on a Chip 2013;13:1070–1078.

[50] B. Bhushan, Nanotribology and nanomechanics of MEMS devices, In: The 9th Annual International Workshop on Micro Electro Mechanical Systems. IEEE; 1996:91–98.

[51] P. Premakumar, A star search for path planning tutorial, http://www.mathworks.com/matlabcentral/fileexchange/26248-a—a-star–search-for-path-planning-tutorial; 2010.

..................Content has been hidden....................

You can't read the all page of ebook, please click here login for view all page.
Reset
18.117.152.250