92 Oscilloscopes
aware of, nor want to know, how much real time elapses
between samples; he or she is only concerned with the timescale
of the reconstructed image. In practice, the sampling rate is
seldom as high as Figures 6.1 and 6.2 might suggest. Typically, the
sampling rate is around 100 ks/s (kilosamples per second). So the
500 MHz triangular wave shown in Figure 6.2 would in fact be
sampled on every 5000th cycle.
To briefly recapitulate. The c.r.t, display is built up of discrete
dots whose vertical positions correspond to the signal voltage at
the time of sampling and whose horizontal positions correspond
to the time delay between the beginning of the next sampled
waveshape and the moment when the sample is taken. To
recognize the beginning of the waveshape, the instrument uses
trigger circuitry much like that of a conventional oscilloscope.
The trigger circuitry will be preceded by a divider stage and
gating, to limit the sampling rate to 100ks/s or so. High speed
logic circuits such as ECL (emitter coupled logic) can cope up to
1 GHz or higher. For triggering from very high-frequency
signals, e.g. from several GHz upwards, the design may employ
a tunnel diode or similar specialized circuitry, providing 'trigger
countdown'.
signal
l I
trigger ~At
! ~ ~ ~
~t ~I
~t i~ a ~J~ ~I ~l
0.2 --~i4 ,nsl o , , , ' i '
real
, I ' , I ' ' I
....
time Im"""~"'lJ"lJl'"Jll"'l'l"l'l'liJ .... 1 .... l~,,,lh,,,l,,,,l,,,jll~,l,l,,,,l,u,,,l .... l,,i,l,,,,l,
0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100105110ns
samples _~__-o- -_ _._
taken ~--4---~ e .... 4---~--- ~ ---i -----'~
oscilloscope display
[ ";'" 1
lee
o%
equivalent
time 0 1 2 ns
Figure 6.2 A sampling oscilloscope may take one sample per cycle of the input
as in Figure 6.1, or, more typically, one sample per n cycles of the input, as here.
In this example n = 5, but in practice n could equal, say, 50, 500, 5000 or any
larger number